Choose
the software driver update from the downloads list.
Select
Save this file to disk.
Click
OK.
Specify
a folder on the hard drive from the list, and then click Save.
Click
Open to extract the file using WinZip, or click Open Folder
for more options.
Recommended: choose to open the folder and click the file with
the right-mouse button, point to WinZip, and then click Extract
to here in the menu.
How to
upgrade software and drivers:
Start HyperTerminal.
Set the port properties
to:
Bits per second: 19200
Data bits: 8
Parity: None
Stop bits: 1
Flow control: Xon/Xoff
Emulation: ANSI
Connect the multiprobe
to the PC using the interface cable.
Start communications
with the multiprobe and the PC. The header information shows up and
then real-time data.
Select Level
3, and then press Enter.
Type the password
Hydlab for Level 3 Security, and then press Enter.
The default password is case sensitive. The Main Menu returns and
the status line at the bottom of the screen reflects the unit is logged
at Level-3 security
Note: If the default password fails for Level-3 security, then
it has been changed. Contact your organization for the new password.
Select Software
from the multiprobe Main Menu, and then press Enter.
Choose option
Sensor or Main.
The option Main is used to load the upgrade of the Main MPL Motherboard
software
The option Sensor is used to load the upgrade of the sensor driver.
If you select Main, HyperTerminal displays a message indicating the
Sonde is waiting for XMODEM-1K file transmission. A string of letters
"C" displays when the Sonde is communicating with the PC.
Proceed to step 10.
If you select Sensor, select the multiprobe parameter to update the
sensor driver, and then Enter.
Select Add-Reset
from the three options, and then press Enter.
HyperTerminal shows a message indicating the Sonde is waiting for
XMODEM-1K file transmission. A string of letters "C" displays
when the Sonde is communicating with the PC.
Select
Transfer in the HyperTerminal menu.
Select Send
File.
Use the Browse to locate the unzipped driver file in the HyperTerminal
pop-up.
Choose Protocol
to 1K-XMODEM from the drop-down menu.
Click Send
to start the file transfer.
Note: The file transfer will fail due to a time out. Press any key
to restart communications between the multiprobe and PC. Return
to step 10 and proceed.
A message will display when the transfer is completed. The multiprobe
restarts and resets Level-2 security.
Note: If Main MPL Motherboard software was installed, time and temperature
parameters are displayed. You can add other parameters using Setup,
Display, and Parameters menus.
After installing upgraded MPL Motherboard software: SDI 12 defaults
are disabled. Enable SDI 12 defaults using Setup, I/O, and SDI Address
menus. Type the the SDI address to enable SDI
12. Type the capital letter X to disable SDI 12.
After installing upgraded sensor drivers: An asterisk is added to
the parameter next to the installed driver. The asterisk indicates
calibration values have been reset to default parameters. Recalibrating
the parameters removes the asterisk.
Select HL
and press Enter to confirm file transfer from PC to multiprobe.
Select Main
and press Enter to confirm software revision.
Select Sensors
and press Enter to confirm driver revision. Revision levels are shown
with associated parameters.