OTT LogoSens DataLogger

Supply voltage +8 to +16 V DC; typ. +12 V
Power consumption (at 12 V)  
          active, taking measurements approx. 50 mA (max. 1 min. per sensor and measurement interval)
          active, no measurements approx. 30 mA
          sleep, special inputs active approx. 1.5 mA
          sleep, clock only mode approx. 0.4 mA
Data storage 1 MB (optional 4 MB) for approx. 100,000 measured values; non-volatile ring memory (no loss of data in case of voltage failure)
Over-voltage protection integrated; voltage life of all inputs up to 36 V DC; all inputs and outputs are EMV protected to valid norms
Display graphic design dot matrix 122 x 32 pixel
   
Inputs  
Number of physical input channels 8; 16 with expansion board
Pulse inputs (terminal G and H)  
          frequency (max. 50 Hz)
          debouncing 0 ms
          pulse time 10 ms
Voltage inputs (5 V /10 V)  
          Ri for GND-related voltages 4 MOhm
          Ri for floating voltages > 100 MOhm
          accuracy 0.1 % of scale max.
          resolution 1 mV
Current input (0(4) … 20mA)  
          Ri typ. 400 Ohm
          accuracy 0.1 % scale max.
          resolution 1 µA
Pt 100  
          accuracy +/- 0.1 K (without cable)
          resolution 0.01 K
          measuring range -30 °C … +85 °C
Digital input RS232-*; RS485-*; SDI-12-interface (* OTT protocol)
Max. current 25 mA (supply of sensors)
   
Outputs
Current output 4 to 20 mA
Relay outputs bounce-free  
          Max. current 800 mA
  electronic short circuit protection with automatic identification
Voltage outputs
          +5 V 1 A
          Vbat looped voltage supply
          -12 V 100 mA
   
Dimensions L x W x H 8.5 x 5.6 x 1.9 in.
Weight approx. 1 lb.
Housing made of ABS
Protection IP 30
Temperature range -22 °F to 158 °F
Acceptable air humidity 10 to 90 %; no condensation


Hach Company Privacy Policy Site Map